This important test measures the influence of a moving marker on a static marker, which
remains immobile, or on another marker. Theoretically, movement analysis should
minimize noise created by movements which do not actually exist. Slide 19,
illustrates the moving markers relative to the stationary marker.
Slide 19. Measures from Static Marker.

The
stationary test measured the ability of the system to maintain static measurements when a
moving marker passed in the vicinity of a stationary marker. With most marker-based
systems, this is a major problem which necessitates human intervention to manually correct
the problem. The point is, a stationary marker should not move.
Stationary Test 0-cm
|
|
Rank |
Max |
Rank |
Markers |
Rank |
Total |
Rank |
Sub-Total |
|
0 cm |
Score |
Error |
Score |
Crossed |
Score |
Error |
Score |
Score |
Ariel |
0.140 |
5 |
0.245 |
5 |
No |
5 |
0.385 |
5 |
20 |
Motion |
0.147 |
4 |
0.604 |
4 |
No |
5 |
0.751 |
4 |
17 |
Peak *** |
0.282 |
2 |
1.183 |
2 |
Yes |
0 |
1.465 |
3 |
7 |
Qualisys |
0.299 |
1 |
3.206 |
1 |
No |
5 |
3.505 |
1 |
8 |
Vicon *** |
0.183 |
3 |
1.725 |
3 |
Yes |
0 |
1.908 |
2 |
8 |
***Markers crossed,
only subset of data which was properly indentified was use for analysis.
Table 4. 0-cm test results
Table 4
illustrates the data for the 0-cm separation. As can be seen, the APAS system
performed the best with an average error amounting to 1.4 mm in a 3000 mm volume.
This result was the lowest of all other companies. The maximum error amounted to 2.4
mm, again the lowest of all companies. For both Peak and Vicon, only a subset of the
data points were used since the markers crossed and switched.

Slide 20. RMS Error at 0-cm separation.
Slide 20
illustrates the error for the 0-cm separation. Motion Analysis performed the best
and the APAS was second. Peak and Vicon were unable to keep the markers on track so
there was switching of the markers. However, the investigator of this research
chose not to include these errors in the statistical analysis. If these measurements
had been included in the statistics, the error would be much larger !!!
2-cm-separation
|
|
Rank |
Max |
Rank |
Markers |
Rank |
Total |
Rank |
Sub-Total |
|
2 cm |
Score |
Error |
Score |
Crossed |
Score |
Error |
Score |
Score |
Ariel |
0.210 |
2 |
0.588 |
2 |
No |
5 |
0.798 |
2 |
11 |
Motion |
0.153 |
3 |
0.571 |
3 |
No |
5 |
0.724 |
3 |
14 |
Peak *** |
0.256 |
1 |
1.623 |
1 |
Yes |
0 |
1.879 |
1 |
3 |
Qualisys |
0.054 |
4 |
0.210 |
5 |
No |
5 |
0.264 |
5 |
19 |
Vicon *** |
0.083 |
5 |
0.458 |
4 |
Yes |
0 |
0.541 |
4 |
13 |
Table 5. 2-cm separation test results
INCREDIBLE !!! In the 2 cm separation
test, Qualisys had the smallest total error. Vicon and Peak again had a problem with
switching of the points between the stationary and the moving targets. However, in
spite of this crossing problem, they were given a score. One should keep in mind
that the error is only 2 mm in 3M volume. This is approximately a .06 percent error.
Slide 21. Maximum Error at 0-cm Separation
As can be
seen in Slide 21, the APAS system out performed all of the other systems. As was
seen previously with other conditions, both Peak and Vicon demonstrated problems with a
switching of the targets although the largest error was observed in the Qualisys system.
Slide 22. Marker Switching at 1-cm Separation.

Slide 22
shows the incredible error the Peak and Vicon produced in tracing markers. This means that
when markers approach approximately 1 cm distance from each other, neither the Peak nor
the Vicon system can detect the difference between the markers. The APAS was able to
do a perfect job. Markers were never switched regardless of the distance between
them. Even at a 0-cm separation, which means no separation, the APAS system
was able to detect the markers. |